B Medical Systems Sarl RCW12

Passive Vaccine Transport Box LR 7L

Product owner: B Medical Systems

Sub-category
Manufacturer
B Medical Systems Sarl
PQS No
E004/004
First pre-qualification date
22-03-2010

Title Author Year Type Language Shared by
Decommissioning and safe disposal of cold chain equipment UNICEF/WHO 2018 Report English Moderator
B Medical System brochure B Medical Systems 2018 Brochure English Moderator
Cold life at +43°C
114,9 hours
Cool life at +43°C
26,4 hours
Warm life at - 20°C
40,9 hours
Coolant packs
0.6 L
Vaccine storage dimensions (cm)
25,1 x 17,6 x 20,9
Gross storage volume (Litre)
9,23 Litres
Vaccine storage capacity (Litre)
7
Model coolant pack
0,6 L
Insulation thickness (mm)
90 - 105 mm
Internal dimensions (cm)
34 x 26 x 27
Number coolant packs
14 units
Pieces per package
1 unit
Shipping volume (m3)
0,148
Shipping weight per carton(kg)
17 kg
Weight empty(Kg)
11,7
Weight fully loaded (Kg)
23,3
Minimum Rated Ambient T°
N/A
Mfr unique reference
RCW12
External dimensions
55 x 47,5 x 49,9
Quality standard
ISO 9001:2008;#ISO 13485;#Other

Above Zero: Strategies to Prevent Vaccine Freezing

This 4-minute animated video emphasizes the difficulty and importance of maintaining proper temperatures in the vaccine cold chain. The risk of exposing vaccines to freezing temperatures are now known to be significant. These risks can be reduced through health worker training, temperature monitoring, investing in new freeze-safe cold chain equipment, and in vaccine formulations that can withstand cooler temperatures. For more information: World Health Organization | http://www.who.int/immunization/programmes_systems/supply_chain/optimize/en/index.html PATH | http://www.path.org TECHNET21 | http://www.technet-21.org View the video in French: http://youtu.be/UEGMIadQeYI.


How to Create an Equipment Inventory

You will need a contingency plan for when things go wrong with your equipment. An inventory list helps technicians plan repairs, manage spare parts, and see which locations may need new or additional equipment. Learn how to fill out an equipment inventory, step-by-step. This EPI training video provides knowledge and skills based on WHO guidance (for example, Mid-Level Manager and Immunization in Practice modules) to support routine immunization in the field. Find more at www.immunizationacademy.com.


Cold Chain Equipment Production at B Medical Systems

At B Medical Systems, we use the rotational moulding technique to manufacture our vaccine cold chain equipment. The process guarantees products of extreme robustness, high quality and zero corrosion.